Bench Testing of the 8 N Anti-Slippage Specification at the Handle–Shaft Locking Interface of Multi-Articulating Laparoscopic Instruments

Corresponding author: Hongchang Ma, mahc860428@gmail.com
DOI: 10.12201/bmr.202610.00026
Statement: This article is a preprint and has not been peer-reviewed. It reports new research that has yet to be evaluated and so should not be used to guide clinical practice.
  •  

    Abstract: Background: Multi-articulating laparoscopic instruments require a reliable mechanical connection between the handle and shaft to maintain the intended instrument configuration during surgical procedures. The handle–shaft locking interface should resist unintended relative displacement under external loading. However, the appropriateness of a specific anti-slip threshold must be evaluated in the context of the instrument design and its intended conditions of use. Objective: To evaluate the mechanical performance of an 8 N anti-slip specification for the handle–shaft interface of multi-articulating laparoscopic instruments in the locked state using a controlled bench-top model. Methods: A bench-top mechanical evaluation method was developed using the test instruments, a mechanical fixation fixture, a calibrated force gauge, and displacement measurement equipment. After engagement of the handle–shaft locking mechanism, external forces were applied to assess its resistance to relative displacement. Test procedures included axial loading, loading in relevant transverse directions, and repeated locking–unlocking cycles. The primary outcome measure was the force at the onset of relative slippage at the locking interface. Secondary outcome measures included displacement under load, complete locking failure, and changes in mechanical performance following repeated operation. Acceptance criteria should be predefined before confirmatory testing. Results: All 30 instruments from three batches met the 8 N axial requirement (30/30). The axial slip-onset force was 12.40 ± 0.97 N (range: 10.80–14.00 N). Under a sustained load of 8 N for 60 s, the mean relative displacement increased from 0.06 mm to 0.10 mm, with a mean residual displacement of 0.02 mm. After 1,000 cycles, the axial slip-onset force was 12.00 ± 0.97 N (range: 10.40–13.60 N). Conclusion: This study provides a bench-testing framework for evaluating the 8 N anti-slip specification at the handle–shaft locking interface of multi-articulating laparoscopic instruments. Whether this specification provides adequate mechanical stability can only be determined by comparing the measured performance against engineering acceptance criteria predefined on the basis of the instrument's intended use and foreseeable loading conditions.

    Key words: Multi-articulating laparoscopic instruments; handle–shaft interface; locking mechanism; anti-slip resistance; mechanical verification; bench testing; minimally invasive surgery.

    Submit time: 28 October 2026

    Copyright: The copyright holder for this preprint is the author/funder, who has granted biomedRxiv a license to display the preprint in perpetuity.
  • 图表

  • Chen Xin Yan, Guo Fang, Li Wei, Chen Hai Tang. Efficacy of the single-port laparoscopic technique in the treatment of ovarian cysts. 2024. doi: 10.12201/bmr.202410.00071

    ZHOU Qiaoyun, HUO Jing, JING Shengyang, ZHANG Xin, XIA Enlan, HUANG Xiaowu. Application of Hysteroscopic and Laparoscopic Surgical Videos in Gynecological Clerkship Teaching. 2025. doi: 10.12201/bmr.202505.00037

    刘高新丽, Zhang Jiacheng, Zhao Siyuan, Li Huanan, Wang Jingui. Progress of the study on the treatment of gastrointestinal dysfunction after laparoscopy by external treatment of Chinese medicine. 2024. doi: 10.12201/bmr.202412.00040

    KE Qingxian, HUANG Limin, LIU Xiaoying, LIN Lizhen. Effect of health education based on LEARNS model on urinary incontinence patients after laparoscopic radical prostatectomy. 2024. doi: 10.12201/bmr.202409.00038

    张莹, Zhangfeng, qiaolin, zhangyuelin, houzhongchen. Analysis on the Advantages of Transumbilical Single-Port Laparoscopy in the Treatment of Pregnancy Complicated with Ovarian Teratoma. 2026. doi: 10.12201/bmr.202604.00179

    Jiaxing Maternal and Child  Jiaxing, Han qin Li. CaiI Xiaoji, Wu Xiaojie *, LI Hanqin. 2025. doi: 10.12201/bmr.202501.00007

    Kang ziming, Shang panfeng. . 2024. doi: 10.12201/bmr.202409.00018

    沈卓, zhaoying, GaoTao. Research advances of natural killer cells at the maternal-fetal interface. 2025. doi: 10.12201/bmr.202505.00023

    chenjunjie, hudonglai. . 2025. doi: 10.12201/bmr.202507.00039

    李威, Geng Zhong li, Angshaer·Sulaiman. Innovation and evolution of neural monitoring technology in endoscopic thyroid cancer surgery. 2024. doi: 10.12201/bmr.202411.00051

  • ID Submit time Number Download
    1 2026-10-12

    10.12201/bmr.202610.00026V1

    Download
  • Public  Anonymous  To author only

Get Citation

Hongchang Ma. Bench Testing of the 8 N Anti-Slippage Specification at the Handle–Shaft Locking Interface of Multi-Articulating Laparoscopic Instruments. 2026. biomedRxiv.202610.00026

Article Metrics

  • Read: 72
  • Download: 1
  • Comment: 0

Email This Article

User name:
Email:*请输入正确邮箱
Code:*验证码错误